Smart Plus (PHT-35LHS)
Technical Manual
VT-TECH-Smart Plus-001
183 / 188
COPYRIGHT© 2018 VATECH
.
ALL RIGHTS RESERVED
3. Error Message List
Tube Error Message Sent From MCU To PC
Command
Description
[SMP_ERRR_0001]
The tube is not ready
[SMP_ERRR_0002]
Tube Mono Block and Inverter Board Cable are not connected
[SMP_ERRR_0003]
Current is more than maximum permissible current while Inverter
Board is working (X-Ray exposure)
[SMP_ERRR_0008]
The temperature of Mono-Block is over 55
℃
[SMP_ERRR_0009]
The current of High Voltage (EP, IP) is more than permissible
current during X-Ray exposure.
[SMP_ERRR_0010]
Although X-Ray Switch is turned on, X-Ray On Command is not
entered in the System
[SMP_ERRR_0011]
Although X-Ray Switch is turned off, X-Ray Off Command is not
entered in the System
[SMP_ERRR_0012]
During X-Ray exposure, kVp Feedback is less than -20 kVp off from
the setting value.
Error “3” is displayed on Inverter Board.
[SMP_ERRR_0013]
During X-Ray exposure, kVp Feedback is more than +20 kVp off
from the setting value.
Error “4” is displayed on Inverter Board.
[SMP_ERRR_0014]
During X-Ray exposure, mA Feedback is less than 50% off of the
setting value.
Error “5” is displayed on Inverter Board.
[SMP_ERRR_0015]
During X-Ray exposure, mA Feedback is more than 150% off of the
setting value.
Error “6” is displayed on Inverter Board.