Smart Plus (PHT-35LHS)
Technical Manual
VT-TECH-Smart Plus-001
39 / 188
COPYRIGHT© 2018 VATECH
.
ALL RIGHTS RESERVED
2. CBCT Geometry Calibration
2.1. Recovering Factory Default Calibration Data for CBCT Geometry
Calibration
Follow the procedure below to recover the factory default calibration data.
If you installed a new sensor, do not use the factory default sensor calibration data.
Instead, go to 1.1 CBCT Preliminary Sensor Calibration and follow the instructions.
1)
Go to
C:\VCaptureSW\Acquisition\CBCT\WidePano\
2) Rename the CAL folder to
CAL_BAKCUP
or something similar.
3) Find the
CAL_GEO
folder and rename it to CAL.
2.2. Setting up Bead Phantom Offset Value
Sensor Leveling
If you installed a new sensor, make sure that you made it level before CBCT
Geometry Calibration.
1) Prepare a tripod and put it on the floor near the equipment.