Smart Plus (PHT-35LHS)
Technical Manual
VT-TECH-Smart Plus-001
71 / 188
COPYRIGHT© 2018 VATECH
.
ALL RIGHTS RESERVED
8) If the standard is met, do the steps that follow and do the steps 1) through 8) again.
To adjust the X value,
①
Type [SPM_AX?_] and click Send to read the current X-axis position.
②
Type [SPM_AXST_####] and click Send to adjust X axis position.
To adjust the Y value,
①
Type [SPM_AY?_] and click Send to read the current Y-axis position.
②
Type [SPM_AYST_####] and click Send to adjust Y-axis position.
•
When you change AXST, AYST value by 10, it will be changed by approximately
5 pixels.
•
To increase X value, decrease AXST value.
•
To increase Y value, increase AYST value.
To make the cross-line level,
①
Open reconstruction.vxm.
②
Change STARTANGEL value (default: 0.0) under Geometry section.
ex) To tilt the angle by 1° clockwise, type 1,
To tilt the angle by 1° counterclockwise, type 359.
③
Save reconstruction.vxm file and exit.