Smart Plus (PHT-35LHS)
Technical Manual
VT-TECH-Smart Plus-001
85 / 188
COPYRIGHT© 2018 VATECH
.
ALL RIGHTS RESERVED
13) Click CBCT tab, type average value in each box for the FOV1 under HIGH RESOLUTION.
Click
Save
to keep the changes, then click
Close
.
Do not measure the Metal value and use the fixed value of 2300 for every FOV.
14) Do steps 2) through 14) for other modes as follows:
Count
FOV
Image Quality
Voxel
Gender
Tooth type
100 x 85(FOV.1)
Green
Standard (0.2)
Man
Hard
100 x 70(FOV.2)
High Resolution
Standard (0.2)
Man
Hard
100 x 70(FOV.2)
Green
Standard (0.2)
Man
Hard
Summary of Contents for Smart Plus PHT-35LHS
Page 1: ...Technical manual Model PHT 35LHS Version 1 16 English...
Page 2: ......
Page 190: ...www vatech com...