The instrument is radio and television interference suppressed according to DIN EN 60601-1-2.
Nevertheless, electromagnetic interference can under certain conditions impair the device function.
Interference can, for example, be caused by mobile telephones. If you detect such interference, switch
off the AKITA JET and increase the distance from possible sources of interference (see also section 11.6).
11.4
Storage- and Transport Conditions
Ambient conditions for transport and storage of the AKITA JET Inhalation System.
•
Ambient temperature:
-25 to 70°C
•
Relative humidity: 15 to 93% non-condensing atmospheric humidity
•
During storage and transportation, the equipment should not be exposed to constant direct
solar radiation.
•
Use the carrying handle provided when transporting the unit.
•
Do not store the nebuliser set, control unit or power supply in damp areas (e.g. bathroom) or
transport them together with damp objects.
•
In the event of extreme fluctuations of the ambient temperature, the system can malfunction
due to condensation.
•
In general between treatments the nebuliser set should be stored as described in section 5.4.
•
After transport or storage of the device: Wait one hour before treatment for temperature
acclimatization.
11.5
Service Life
The AKITA JET Inhalation System is intended for repeated use. The average life of the components is
given in the table below.
C
OMPONENT
S
ERVICE
L
IFE
AKITA JET control unit (the device can continue
to be used after maintenance)
10 Years
(at a daily inhalation treatment time of 20
minutes)
AKITA JET nebuliser set
12 Months
(maximum 300 sterilisation cycles)
Filter
12 Months