C-6
Stress Measurement
Figure C-6: Display Parameters Dialog Box: Type Tab
You can select the following types of stress plots from the dialog box:
Curve Fit
Curve fit plots in cyan.
PreDeposition
Predeposition data curve plots in blue.
NOTE –
This choice is available only if a pre-deposition data file was loaded.
PostDeposition
Post-deposition data curve plots in green.
Film Stress Curve
Film stress curve plots in orange.
NOTE –
This choice is relevant only if a pre-deposition data file was loaded.
Constraints and Limitations
When using the stress measurement options you must:
•
Manually level the stage and verify that the stage is in the same leveled position before running
both the pre- and post-stress scans.