4-4
Single Scan Operation
Figure 4-3: Scan Routines Window with Default Scan Routine
You can edit the options for this automation program and add or modify scan routine parameters (see
Save the automation program according to the instructions in
Saving an Automation Program on
.
Defining Scan Location and Length
This procedure entails determining the appropriate scan starting location and length for measuring
the sample.
1
Select
Window
>
Sample Positioning
from the menu to display the
Sample Positioning
window.
2
Select
Profiler
>
Tower Down
.
3
Use the
Illumination Adjustment
icons to adjust the illumination to view the sample.
4
With the stylus tower down, initiate Stage Tracking (left-click in the camera view pane) for fine sample
positioning. The image should now track the motion of the mouse.
5
With the stylus reticle properly positioned on the sample, click the mouse button a second time to
deactivate Stage Tracking. This is your starting point.
6
Select
Edit
>
Define Scan Location/Length
. The
Location/Length for Routine #: 1 of 1
dialog box
appears (see
X
,
Y
, and
T
fields display the current X-Y stage position.
Summary of Contents for Dektak 150
Page 1: ......
Page 7: ......
Page 17: ......
Page 81: ......
Page 105: ......
Page 149: ......
Page 191: ......
Page 251: ......
Page 257: ...A 6 Facilities Specifications Figure A 6 Dektak 150 without Enclosure Side View ...
Page 258: ...Facilities Specifications A 7 Figure A 7 Dektak 150 Dimensions without Enclosure Top View ...
Page 259: ...A 8 Facilities Specifications Figure A 8 Dektak 150 without Enclosure Center of Mass ...
Page 273: ......
Page 283: ......
Page 321: ......
Page 331: ......
Page 333: ......
Page 336: ......
Page 337: ......
Page 349: ......