7-6
Scan Routine Parameters
Figure 7-6: Scan Parameters Dialog Box: Length
The scan length can also be set from the
Sample Positioning
window by selecting
Edit
>
Enter
Scan Length
. You can enter the scan length manually using Stage Tracking, as discussed in
.
Scan Duration/Speed
The
Duration
setting displays the amount of time it takes to complete a given scan. Scan duration, in
conjunction with scan length, determines the horizontal resolution of a scan. Therefore, scan speed is
directly related to the resolution.
For example, a 13-second scan provides 3900 sample data points. You can set the scan duration from
3 to 200 seconds for a maximum of 60,000 data points per scan.
Select a longer scan duration for long scan applications and measurements of very fine surface
roughness requiring the highest horizontal resolution. When high throughput is the primary
consideration, use a shorter scan duration. For most applications, a 10-20 second scan provides
adequate resolution and throughput.
1
Click any parameter in the
Scan Parameters
section of the
Scan Routines
window (see
display the
Nominal Parameters
tab of the
Scan Parameters
dialog box.
2
In the
Duration
field, enter the desired scan duration (in seconds) using the keyboard (see
3
Click
OK
to close the dialog box and enter the duration into the scan program.
Summary of Contents for Dektak 150
Page 1: ......
Page 7: ......
Page 17: ......
Page 81: ......
Page 105: ......
Page 149: ......
Page 191: ......
Page 251: ......
Page 257: ...A 6 Facilities Specifications Figure A 6 Dektak 150 without Enclosure Side View ...
Page 258: ...Facilities Specifications A 7 Figure A 7 Dektak 150 Dimensions without Enclosure Top View ...
Page 259: ...A 8 Facilities Specifications Figure A 8 Dektak 150 without Enclosure Center of Mass ...
Page 273: ......
Page 283: ......
Page 321: ......
Page 331: ......
Page 333: ......
Page 336: ......
Page 337: ......
Page 349: ......