Scan Routine Parameters
7-7
Figure 7-7: Scan Parameters Dialog Box: Duration
Scan Resolution
The
Resolution
parameter (see
) displays the horizontal resolution for the scan length and
scan speed (duration) entered into the scan routine. The scan resolution is expressed in
μ
m/sample,
indicating the horizontal distance between data points. Data points are the points along the scan path
at which data samples are taken. The more data points taken during a given scan length, the shorter
the distance between data samples. Therefore, a scan routine with the lowest number of µm per
sample provides the best possible horizontal resolution.
The Dektak 150 provides horizontal resolution with a maximum 60,000 data points available per
scan. Scan length and scan duration determine the horizontal resolution of the Dektak 150. The
profiler maintains a constant sampling rate of 300 data points per second. By slowing scan speed, you
can process more samples with a given scan length over a longer period of time. Scan duration may
be set anywhere from 3 to 200 seconds. The examples below provide the number of data points per
scan at various scan durations for a 2000 µm scan length.
Summary of Contents for Dektak 150
Page 1: ......
Page 7: ......
Page 17: ......
Page 81: ......
Page 105: ......
Page 149: ......
Page 191: ......
Page 251: ......
Page 257: ...A 6 Facilities Specifications Figure A 6 Dektak 150 without Enclosure Side View ...
Page 258: ...Facilities Specifications A 7 Figure A 7 Dektak 150 Dimensions without Enclosure Top View ...
Page 259: ...A 8 Facilities Specifications Figure A 8 Dektak 150 without Enclosure Center of Mass ...
Page 273: ......
Page 283: ......
Page 321: ......
Page 331: ......
Page 333: ......
Page 336: ......
Page 337: ......
Page 349: ......