Scan Routine Parameters
1
Click any parameter in the
Scan Parameters
section of the
Scan Routines
window (see
display the
Nominal Parameters
tab of the
Scan Parameters
dialog box.
2
In the
Stylus Force
field, enter the desired stylus force (see
3
Click
OK
to close the dialog box and enter the stylus force into the scan program.
NOTE –
If you system includes the
N
-Lite option (see
), that option engages
automatically when you enter a Stylus Force value that is < 3 mg.
Figure 7-10: Scan Parameters Dialog Box: Stylus Force
Measurement Range
The available vertical resolution depends upon the
Measurement Range
selected. When measuring
extremely fine geometries, the 65 kÅ range provides a vertical bit resolution of 1 Å. For general
applications, the 10 Å vertical resolution of the 655 kÅ range is usually adequate. When measuring
thick films or very rough or curved samples, select the 5240 kÅ range with 80 Å resolution.
1
Click any parameter in the
Scan Parameters
section of the
Scan Routines
window (see
display the
Nominal Parameters
tab of the
Scan Parameters
dialog box.
2
In the
Meas
Range
field, select one of the measurement ranges from the drop-down list:
65 kÅ
,
655 kÅ
or
5340 kÅ
(or
1 mm
, optional). See
3
Click
OK
to close the dialog box and enter the measurement range into the scan program.
Summary of Contents for Dektak 150
Page 1: ......
Page 7: ......
Page 17: ......
Page 81: ......
Page 105: ......
Page 149: ......
Page 191: ......
Page 251: ......
Page 257: ...A 6 Facilities Specifications Figure A 6 Dektak 150 without Enclosure Side View ...
Page 258: ...Facilities Specifications A 7 Figure A 7 Dektak 150 Dimensions without Enclosure Top View ...
Page 259: ...A 8 Facilities Specifications Figure A 8 Dektak 150 without Enclosure Center of Mass ...
Page 273: ......
Page 283: ......
Page 321: ......
Page 331: ......
Page 333: ......
Page 336: ......
Page 337: ......
Page 349: ......