Calibration and Maintenance
9-11
f.
Carefully lower the stylus exchange tool. The stylus is held in place magnetically.
Figure 9-13: Fixture with Captured Stylus
4
To remove the stylus from the stylus exchange tool:
a.
Disengage the magnet by rotating the wheel in either direction so a notch is visible.
b.
Gently lift the stylus out of the tool, and then place it in its protective case.
Installing a New Stylus
After you have removed an old stylus, follow these steps to install a new one:
1
Locate the new stylus and stylus exchange tool. Each stylus comes in its own protective plastic case.
2
The stylus exchange tool is designed to magnetically hold the stylus until it is seated in the sensor head.
Turn the thumb wheel on the exchange tool so that the green dot (see
) is NOT visible (that is,
the magnet is disengaged).
3
Open the stylus case, and gently remove the stylus. Lift the stylus by its central disk, not by the beam.
4
Place the stylus into the exchange tool. Align the round disk with the silver magnet on the tool. The stylus
arm must be centered within the long trench (see
). The stylus tip should point down.
Figure 9-14: Alignment of Stylus and Stylus Exchange Tool
Stylus
Stylus misaligned with exchange tool
Stylus correctly aligned with exchange tool
Green dot
Incorrect
position
Summary of Contents for Dektak 150
Page 1: ......
Page 7: ......
Page 17: ......
Page 81: ......
Page 105: ......
Page 149: ......
Page 191: ......
Page 251: ......
Page 257: ...A 6 Facilities Specifications Figure A 6 Dektak 150 without Enclosure Side View ...
Page 258: ...Facilities Specifications A 7 Figure A 7 Dektak 150 Dimensions without Enclosure Top View ...
Page 259: ...A 8 Facilities Specifications Figure A 8 Dektak 150 without Enclosure Center of Mass ...
Page 273: ......
Page 283: ......
Page 321: ......
Page 331: ......
Page 333: ......
Page 336: ......
Page 337: ......
Page 349: ......