Dektak 150 System Overview
1-7
Measurement Output
The
Data Plot
window shows the live scan data, the user-set scan parameters, and the results of the
calculations of any user-selected analytical functions, which provide detailed statistical information
about the profile data. For more information, see
Analytical Functions on page 1-7
DEKTAK 150 SOFTWARE FUNCTIONS
The Dektak 150 Version 9 software carries out all of the functions necessary to taking a single- or
multiple-scan measurement and calculating analytical functions. In addition, it performs:
•
MicroForm™ adjustment
, which reveals difficult shapes and overcomes steep slopes, thus
improving the accuracy of the slope calculation. The MicroForm algorithm adjusts the lateral
plot to become a function of the vertical position of the stylus, thus compensating for the arcing
motion of the stylus.
•
Step detection
, which automatically levels, detects and measures single or multiple steps in a
single scan. It also provides an average of all the steps.
The following sections describe some of the major functions of the Dektak 150 software.
Automation Programs
Automation Program files can store a number of scan routines on the hard disk. Scan routines, along
with their stage locations and analytical functions, are inserted into the automation program.
Automation Programs are stored for various applications in Windows file format on the hard disk,
giving the Dektak 150 virtually unlimited program storage capability.
The Automation Program Summary (APS)
Report provides data in tabular form on the just-
concluded or saved automation program. In addition to showing basic information such as the
number of scan routines, this window displays the cursor locations and statistical information
regarding each analytical function (see
Analytical Functions on page 1-7
).
Scan Routines
A Dektak 150 scan routine consists of up to sixteen individual scan parameters that you can select
using the mouse. You can determine parameters such as scan length and speed, software leveling, and
stylus force. A maximum of 10,000 scan routines can be entered into each automation program file.
Analytical Functions
The Dektak 150 has a wide range of analytical functions available for analysis of roughness,
waviness, step height, and geometrical measurements. You can log up to 30 analytical functions per
scan to the APS.
Summary of Contents for Dektak 150
Page 1: ......
Page 7: ......
Page 17: ......
Page 81: ......
Page 105: ......
Page 149: ......
Page 191: ......
Page 251: ......
Page 257: ...A 6 Facilities Specifications Figure A 6 Dektak 150 without Enclosure Side View ...
Page 258: ...Facilities Specifications A 7 Figure A 7 Dektak 150 Dimensions without Enclosure Top View ...
Page 259: ...A 8 Facilities Specifications Figure A 8 Dektak 150 without Enclosure Center of Mass ...
Page 273: ......
Page 283: ......
Page 321: ......
Page 331: ......
Page 333: ......
Page 336: ......
Page 337: ......
Page 349: ......