9-18
Calibration and Maintenance
Adjusting the Standard Fixed Optics
To adjust the standard fixed optics:
1
Place the vertical standard that came with your system on the stage beneath the stylus, with the printing
facing the front of the system (
). The system scans from front to back, so be sure there are at
least a few millimeters of the standard located to the back of the stylus tip.
2
Press the
Stylus Down
button
to lower the stylus onto the standard (
Figure 9-23: Stylus Lowered onto the Standard
NOTE –
The above activity cannot damage the stylus/sensor. However, clipped data may
appear on the monitor.
3
Adjust the focus lock screw near the base of the optical assembly (see
a.
Loosen the focus lock screw if necessary.
b.
Watching the computer monitor, turn the focus ring until the stylus tip on screen is as sharp as possible
(see
and
).
c.
Finger-tighten the focus lock screw.
Figure 9-24: AdjustIng Focus