Chapter 7: Other Modes of Operation
Local Restricted
This mode of operation is limited to conditions recalled from memory. The
user may not change any parameter, and the Adjust knob is nonfunctional.
All switches are active, except the MODULE and FUNCTION switches.
The Local Restricted mode of operation is useful when limited user interven-
tion is needed. Parameters must be stored in the memory locations in
advance.
1. Select the chosen operating conditions in Manual or Local mode.
2. Select the LOC RES mode with the MODE switch.
3. Turn the KEYLOCK switch to the disable position and remove the key.
4. The user may turn the SOURCE or BEAM switches on or off and change
from one memory location to another.
5. The MPS-3000 CE power supply will function as if it were in Local
mode.
Remote Operation
The MPS-3000 CE power supply must be controlled remotely, by a com-
puter, through an EIA-232 interface.
The only front panel switches that are functional in the Remote mode are
the MODE, SOURCE, and BEAM OFF switches.