50
Command Name: AUTOMATIC DATA LOGGING
Syntax: AD[N]<cr>
Units: N/A
Range: n = 0 or 1
Description: This command is used to enable (AD1) or disable (AD0) auto-
matic data logging.
When enabled, the power supply will transmit its actual running condi-
tions at the time interval specified by the T<n> command.
Logging will begin when the source is turned on and will end when the
source is turned off.
Disable auto data logging using the following command: AD0<cr>
Command Name: AUTOMATIC ERROR LOGGING
Syntax: AE[n]<cr>
Units: N/A
Range: n = 0 or 1
Description: This command is used to enable (AE1) or disable (AE0) auto-
matic error logging.
When enabled, the power supply will transmit any error conditions that
occur while the source and beam are on.
Enable auto error logging using the following command: AE1<cr>
Command Name: SET ACCELERATOR VOLTAGE
Syntax: AV[n]<cr>
Units: volts
Range: n = 0 - 1023
Description: Sets the target value for the accelerator voltage. When the
beam is on, the accelerator voltage will ramp to this value.
Set the accelerator voltage to 400V using the following command:
AV400<cr>
EXAMPLE 8.3 Disable
Auto Data Logging
EXAMPLE 8.4 Enable
Auto Error Logging
EXAMPLE 8.5 Set the
Accelerator Voltage
Summary of Contents for MPS-3000 CE
Page 1: ...MPS 3000 CE Power Supply Technical Manual 425964...
Page 2: ...MPS 3000 CE Power Supply Technical Manual...
Page 90: ...86 Filament Cathode with Filament Neutralizer or PBN Option...
Page 91: ...87 Hollow Cathode and HCN 7 pin or 2 x 4 pin feedthrough...
Page 92: ...88 Hollow Cathode and Hollow Cathode Neut 6 pin feedthrough...
Page 93: ...89 Filament Cathode with Filament Neutralizer or PBN Option...
Page 94: ...90 3cm Source Filament Cathode and Filament Neutralizer...
Page 95: ...91 3cm Source Hollow Cathode and Hollow Cathode Neutralizer...