53
Command Name: SET CATHODE LIMIT
Syntax: CL[n]<cr>
Units: MPS-3000 CE FC/PBN: amperes
MPS-3000 CE HC: volts
Range: MPS-3000 CE FC/PBN: 0.00 - 20.00
MPS-3000 CE HC: 0 - 50
Description:
MPS-3000 CE FC and PBN:
Used to set the cathode filament current’s software limit. The actual fil-
ament current will not ramp above this value.
MPS-3000 CE HC:
Used to set the hollow cathode keeper voltage limit.
Set the MPS-3000 CE FC/PBN cathode filament limit to 5.50A using the
following command: CL5.50<cr>
Set the MPS-3000 CE HC hollow cathode keeper voltage limit to 25V
using the following command: CL25<cr>
Command Name: SET DISCHARGE CURRENT
Syntax: DI[n]<cr>
Units: amperes
Range: n = 0.00 - 6.00
Description: Sets the target value for the discharge current. When the
source is on and in Manual mode, the discharge current will ramp to this
value.
This command is effective only in Manual mode, but is accepted in all
modes and is only valid for an MPS-3000 CE HC power supply.
Set the discharge current to 2.00A using the following command:
DI2.00<cr>
EXAMPLE 8.11 Set the
Cathode Filament Limit
EXAMPLE 8.12 Set the
Keeper Voltage Limit
EXAMPLE 8.13 Set the
Discharge Current
Summary of Contents for MPS-3000 CE
Page 1: ...MPS 3000 CE Power Supply Technical Manual 425964...
Page 2: ...MPS 3000 CE Power Supply Technical Manual...
Page 90: ...86 Filament Cathode with Filament Neutralizer or PBN Option...
Page 91: ...87 Hollow Cathode and HCN 7 pin or 2 x 4 pin feedthrough...
Page 92: ...88 Hollow Cathode and Hollow Cathode Neut 6 pin feedthrough...
Page 93: ...89 Filament Cathode with Filament Neutralizer or PBN Option...
Page 94: ...90 3cm Source Filament Cathode and Filament Neutralizer...
Page 95: ...91 3cm Source Hollow Cathode and Hollow Cathode Neutralizer...