57
Command Name: SET NEUTRALIZER LIMIT
Syntax: NL[n]<cr>
Units: MPS-3000 CE FC: amperes
MPS-3000 CE PBN: amperes
MPS-3000 CE HCN: volts
Range: MPS-3000 CE FC: 0.00 - 15.00
MPS-3000 CE PBN: 0.00 - 10.23
MPS-3000 CE HCN: 0 - 50
Description:
MPS-3000 CE FC/PBN:
Used to set the neutralizer filament currents software limit. The actual
filament current will not ramp above this value.
MPS-3000 CE HC:
Used to set the HCN keeper voltage limit.
Set the MPS-3000 CE FC/PBN neutralizer limit to 5.50A using the follow-
ing command: NL5.50<cr>
Set the MPS-3000 CE HCN keeper voltage limit to 25V using the following
command: NL25<cr>
Command Name: RECALL TARGETS FROM MEMORY
Syntax: R[n]<cr>
Units: N/A
Range: n = 1, 2
Description: Recalls the target parameters from the selected memory. The
recalled values will be the new running parameters.
Recall memory 1 using the following command: R1<cr>
EXAMPLE 8.22 Set
the Neutralizer Limit
EXAMPLE 8.23 Set the
Keeper Voltage Limit
EXAMPLE 8.24 Recall-
ing Memory 1