61
Command Name: REQUEST ALL TARGET PARAMETERS IN COMPRESSED
FORMAT
Syntax: RN[n]<cr>
Units: N/A
Range: n = 1, 2
Response:
<lf><cr>^AAAA,^BBBB,^CCCC,^^^DD,EEEEE,FFFFF,GGGG
G,HHHHH,IIIII,JJJJJ,KKKKK,
<lf><cr><eot>
Where:
Description: Used to request all target values, in the compressed format,
from the selected memory. If the value is not reported for a particular
model, it is replaced with ASCII blank spaces (32).
Request all the target parameters stored in Memory 1 using the following
command: RN1<cr>
Parameters
Model
Units
AAAAA
Beam Voltage
All
Volts
BBBBB
Beam Current
All
mA
CCCCC
Accelerator Voltage
All
Volts
DDDDD
A/B Ratio
All
%
EEEEE
Cathode Filament Limit
HC Keeper Voltage Limit
MPS-3000 CE FC
MSP-3000 CE HC
Amps
Volts
FFFFF
Neutralizer Filament Limit
HCN Keeper Voltage Limit
PBN Filament Limit
MPS-3000 CE FC
MPS-3000 CE HCN
MPS-3000 CE PBN
Amps
Volts
Amps
GGGG
G
Discharge Voltage
Discharge Current
MPS-3000 CE FC
MPS-3000 CE HC
Volts
Amps
HHHHH
Beam Current Tolerance
All
%
IIIII
Discharge Current Threshold
All
Amps
JJJJJ
Cathode Filament Current
MPS-3000 CE FC
Amps
KKKKK
Neutralizer Filament
Current
Emission Current
MSP-3000 CE FC
MPS-3000 CE
HCN/PBN
Amps
mA
EXAMPLE 8.26 Request
all Target Parameters
Summary of Contents for MPS-3000 CE
Page 1: ...MPS 3000 CE Power Supply Technical Manual 425964...
Page 2: ...MPS 3000 CE Power Supply Technical Manual...
Page 90: ...86 Filament Cathode with Filament Neutralizer or PBN Option...
Page 91: ...87 Hollow Cathode and HCN 7 pin or 2 x 4 pin feedthrough...
Page 92: ...88 Hollow Cathode and Hollow Cathode Neut 6 pin feedthrough...
Page 93: ...89 Filament Cathode with Filament Neutralizer or PBN Option...
Page 94: ...90 3cm Source Filament Cathode and Filament Neutralizer...
Page 95: ...91 3cm Source Hollow Cathode and Hollow Cathode Neutralizer...