3
isolation circuit incorporates optical transmitters and photo detectors,
which are coupled together by fiber optic cables. This configuration ensures
that the user accessible EIA-232 and system interlock circuits remain at
Safety Extra Low Voltage (SELV) under normal and single fault conditions.
The customer side supply (+5V DC) is generated from an integral isolating
type DC/DC inverter that inverts the ELV +15V DC supply to the required
SELV +5V DC supply.