To avoid electrical shock, keep clear of “live” circuits. Follow all local lock-out/tag-out procedures before continuing. Disconnect the
controller’s main power before troubleshooting any electrical connection.
WARNING
71
E06
Thermal shutdown.
Fatal error (refer to the NOTE on
page 72
).
Blockage of air flow.
Remove the blockage and cycle the
power.
Defective fan.
Contact
“Service Support” on page 73.
E07
Clock fault.
Defective component.
Contact
“Service Support” on page 73.
E08
Cathode filament or neutralizer filament
error. This is a Fatal error (refer to the NOTE
on
page 72
). If a neutralizer filament is the
cause of this error, SW 1-4 is on (controller
board).
Broken cathode or neutralizer filament.
Replace filament.
Cable is disconnected.
Connect the cable.
E09
Only active if Auto Cathode (AC1 com-
mand). Cathode filament is at its limit and
the discharge current is less than the dis-
charge current threshold (DT command).
Discharge threshold set is elevated (DT command).
Decrease value.
Cathode limit set too low.
Increase value.
Insufficient gas flow.
Raise gas flow.
E13
Beam current is out of the tolerance set by
the BE command.
Insufficient gas flow.
Raise the gas flow.
A large change in beam current was requested.
Disable E13 (BEO) before changing
beam current.
Cathode limit is set too low, 3000 CE FC or PBN only.
Increase limit.
Discharge current set too low.
Increase limit.
E18
HC shorted.
various.
Contact
“Service Support” on page 73.
E19
HCN shorted.
various.
Contact
“Service Support” on page 73.
E20
External interlock failure.
External interlock circuit is incomplete.
Complete the external interlock circuit.
E21
Internal interlock failure. Fatal error (refer to
the NOTE on
page 72
).
Voltage is low from the internal line conditioner.
Cycle the power.
Facility voltage is low.
Raise facility voltage.
Power plug is wired incorrectly.
Rewire power plug.
E22
Cathode filament. Fatal error (refer to the
NOTE on
page 72
); MPS-3000 CE FC or
PBN power supplies only.
Broken cathode filament.
Replace filament.
Cable is disconnected.
Connect cable.
Table 10.2: Error Code Definitions (Continued)
Error
Code Definition
Probable Cause
Suggested Remedy
Summary of Contents for MPS-3000 CE
Page 1: ...MPS 3000 CE Power Supply Technical Manual 425964...
Page 2: ...MPS 3000 CE Power Supply Technical Manual...
Page 90: ...86 Filament Cathode with Filament Neutralizer or PBN Option...
Page 91: ...87 Hollow Cathode and HCN 7 pin or 2 x 4 pin feedthrough...
Page 92: ...88 Hollow Cathode and Hollow Cathode Neut 6 pin feedthrough...
Page 93: ...89 Filament Cathode with Filament Neutralizer or PBN Option...
Page 94: ...90 3cm Source Filament Cathode and Filament Neutralizer...