Packet Tab
632
User-Defined Acceptance Criteria:
•
Accept Seq Errors
- A user-specified threshold value for determining the
acceptable Out of Sequence packet rate for a timed test.
•
Accept Bit Errors
- A user-specified threshold value for determining the
acceptable bit error rate for a timed test.
•
Accept Loss Rate
- A user-specified threshold value for determining the
acceptable Packet Loss rate for a timed test.
Measuring Service Disruption
(Specifically for LAN/WAN Ethernet testing)
The unit can measure Service Disruption switching intervals in your network. These
statistics can be used to determine how long it takes for your network equipment to
react to a service disruption event and switch to a backup path.
When enabled, the trigger event to begin service disruption measurements occurs on
incoming packets with CRC/FCS errors, or when it stops detecting incoming packets.
The duration for every Ethernet service disruption event is measured in milliseconds,
with granular accuracy of .1 ms (100 µs).
When the Action state is configured for Continuous APS, the 10 most recent
measurements are displayed in the History window. Also the Minimum and Maximum
measurement times are logged, as well as an Average duration based on all
measurements taken since the test began.
To measure Ethernet service disruption activity:
1. Select the
Test
tab and choose
Service Disrptn
from the tabs along the left-side
of the screen.
MPA_e_manual_D07-00-129P_RevA00
Summary of Contents for MPA
Page 2: ...MPA_e_manual_D07 00 129P_RevA00...
Page 10: ...MPA_e_manual_D07 00 129P_RevA00...
Page 82: ...MPA_e_manual_D07 00 129P_RevA00...
Page 110: ...MPA_e_manual_D07 00 129P_RevA00...
Page 134: ...MLD Tab 134 100G RS FEC Ethernet 400G RS FEC Ethernet MPA_e_manual_D07 00 129P_RevA00...
Page 255: ...Protocol Tabs 255 MPA_e_manual_D07 00 129P_RevA00...
Page 256: ...OTN Tab 256 MPA_e_manual_D07 00 129P_RevA00...
Page 748: ...MPA_e_manual_D07 00 129P_RevA00...