background image

 

24

 / 

24

 

11.

 

Contact information   

 

 

 

 

 

 

 

 

MedPath GmbH 

Mies-van-der-Rohe-Strasse 8, 80807 Munich, Germany 

 

  0123 

 

 

 

Made by: 

                                            Approved by:

 

 

 

 

Vega Technologies Inc. 

No.505, Shengping Mid.Road, Yangwu, Dalang Town, 523789 
Dongguan City, Guangdong Province, PEOPLE'S REPUBLIC OF CHINA

 

Website: www.vegamedical.com 

 

 

 

Summary of Contents for CN-02MU

Page 1: ...1 24 VEGA TECHNOLOGIES INC INSTRUCTION MANUAL Compressor nebulizer with disposable accessories Model No CN 02MU File Number CN 02MU SPC001 Version Revision No 1 0 Date November 01 2022...

Page 2: ...3 2 Safety precaution 5 3 Information for operation 8 4 Cleaning and maintenance 10 5 Troubleshooting information 11 6 Service information 12 7 Technical specifications 14 8 Symbols 16 9 Disposal 17 1...

Page 3: ...bout the features of this product Any use of this product other than its intended use should always be avoided 1 1 Intended use Intended purpose The compressor nebulizer is intended to aerosolize phys...

Page 4: ...and doctor office and in a room of general household Precautions for use Warnings and cautions described in the instruction manual should be observed 1 2 Product identification 1 3 Disposable accesso...

Page 5: ...ic safety precautions should always be followed including the following Warning Indicates a potentially hazardous situation which if not avoided could result in serious injury Usage 1 For type dose an...

Page 6: ...ieces that can be swallowed 14 Do not use the device while sleeping or sleepy 15 Do not use in anesthetic or ventilator breathing circuits 16 Store the device and disposable accessories in a clean loc...

Page 7: ...the device the main unit may become hot 7 Do not touch the main unit for other than necessary operation such as turning off the power while nebulizing 8 Do not reuse the disposable accessories of comp...

Page 8: ...m the power source Save these instructions for future reference 3 Information for operation 3 1 Functional description The nebulizer is mainly used for the treatment of various respiratory diseases su...

Page 9: ...th the medication prescribed by your physician Fig 2 5 Gently twist the top cover clockwise to re assemble the Nebulizer Bottle Be sure the two sections fit well Fig 3 6 Attach one end of Air Tube to...

Page 10: ...cal outlet 4 Cleaning and maintenance Warning Please make sure the power cord is not plugged into the power outlet when clean and disinfect the product Cleaning It is recommended to clean the Compress...

Page 11: ...the filter cover 5 Troubleshooting information Check the following if your unit should fail during operation You can also refer to the pages of this manual for complete instructions Problem Cause Rem...

Page 12: ...vice Professional shall immediately void this warranty ME EQUIPMENT that is intended for use by a LAY OPERATOR shall indicate that the LAY OPERATOR or LAY RESPONSIBLE ORGANIZATION should contact the M...

Page 13: ...13 24...

Page 14: ...o 5ml Max 6ml Air flow 5 0 to 8 0 Lpm Nebulization rate 0 20 ml min 0 9 salt water solution Operating pressure 8 0 to 13 0 Psi 0 06 to 0 09Mpa Maximum pressure 25 Psi 0 17 Mpa MMAD 5 m Noise level Les...

Page 15: ...CNK M06 3 Air Filter PU CNK S03 4 Nosepiece PP translucent CUK M04 5 Child Mask PVC transparent with mask strap CUK M02 6 Adult Mask PVC transparent with mask strap CUK M01 7 Air Tube PVC 150cm CNK S0...

Page 16: ...16 24 7 3 The particle size distribution curve 8 Symbols 8 1 Symbols and description Refer to instruction manual booklet Manufacturer Batch code Date of manufacture Serial number Caution...

Page 17: ...e handle with care Atmospheric pressure limitation The symbol indicating separate collection for EEE Medical Device 9 Disposal Danger of infection The device and disposable accessories may come into c...

Page 18: ...lp prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF Communications Equipment transmitters and the device as recommended below Warning The device sh...

Page 19: ...intended for use in the electromagnetic environment specified below The customer or the user of the device should assure that it is used in such an environment Immunity test IEC 60601 test level Comp...

Page 20: ...ip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 s Not applicable Mains power quality should be that of a typical commercial or hospital environment If the user of the de...

Page 21: ...nducted RF IEC 61000 4 6 Radiated 3Vrms 150kHz to 80MHz 3V m Not applicable 3V m Portable and mobile RF communications equipment should be used no closer to any part of the device including cables tha...

Page 22: ...strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theo...

Page 23: ...150kHz to 80MHz d 1 2 P 80MHz to 800MHz d 1 2 P 800MHz to 2 5GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output pow...

Page 24: ...bH Mies van der Rohe Strasse 8 80807 Munich Germany 0123 Made by Approved by Vega Technologies Inc No 505 Shengping Mid Road Yangwu Dalang Town 523789 Dongguan City Guangdong Province PEOPLE S REPUBLI...

Reviews: