6 Set up with the display and adjustment module
MINITRAC 31 • 4 … 20 mA/HART - four-wire
40447-EN-130430
Current output
Switching function of the relay
Information:
The simulation is automatically terminated 10 minutes after the last
press of a key.
The sensor calculates a suitable integration time automatically.
Additional settings
In this menu item, the PIN is permanently activated/deactivated. Thus
you protect the sensor data against unauthorized access and unin
-
tended changes. The default setting of the PIN is 0000.
This parameter is described in the operating instructions manual
"
Display and adjustment module
".
In this menu item you can set the actual date and time.
This parameter is described in the operating instructions manual
"
Display and adjustment module
".
When a reset is carried out, all settings (with only a few exceptions)
are reset. The exceptions are: PIN, language, SIL and HART mode.
The following reset functions are available:
Basic adjustments:
Resetting of the parameter adjustmetns to
default values at the time of shipment. Order-specific settings will be
deleted.
Diagnosis/Calculated
damping
Additional adjustments/
PIN
Additional adjustments/
Date time
Additional adjustments
- Reset