4.3 Mounting Orientation
Prior to mounting the SHLD 2, make certain the orientation of the source holder
is correct.
Continuous Level Orientation
For continuous level measurement the source holder must be mounted slightly
above or at the height of the max. level. The radiation must be directed exactly
towards the detector mounted on the opposite side.
The source holder SHLD 2 should be mounted as close as possible to the vessel.
However, with large measuring ranges and small vessel diameters, a gap can
often is not avoidable.
If there are gaps or empty spaces around the installation, provide protective
fences or grids to keep hands away from the dangerous area. Such areas must be
marked accordingly.
Align the source holder according to its beam exit angle
.
Figure 4.1
Continuous Level Orientation
Summary of Contents for SHLD 2
Page 1: ...Operating Instructions SHLD 2 Source Holder Document ID 63138 Radiometric...
Page 2: ......
Page 8: ...6 63138 US 200110 NOTES...
Page 12: ...10 63138 US 200110 NOTES...
Page 26: ...24 63138 US 200110 NOTES...
Page 42: ...40 63138 US 200110 NOTES...
Page 56: ...54 63138 US 200110 NOTES...
Page 68: ...66 63138 US 200110 NOTES...
Page 78: ...76 63138 US 200110 9 2 Dimensions Figure 9 7 Source Holder SHLD 2 Standard Version...
Page 79: ...63138 US 200110 77 Figure 9 8 SHLD 2 Beam Angles Figure 9 9 SHLD 2 Height for Source Exchange...
Page 80: ...78 63138 US 200110 Figure 9 10 Source Holder SHLD 2 with Limit Switch...
Page 81: ...63138 US 200110 79 Figure 9 11 Source Holder SHLD 2 with Interlock Safety Switch...
Page 82: ...80 63138 US 200110 Figure 9 12 Source Holder SHLD 2 with Air Actuator...
Page 83: ...63138 US 200110 81 Figure 9 13 Source Holder SHLD 2 with Air Actuator and Limit Switch...
Page 84: ...82 63138 US 200110 Figure 9 14 Source Holder SHLD 2 Explosion Proof with Limit Switch...
Page 86: ...84 63138 US 200110 NOTES...
Page 87: ......