63138-US-200110
29
Point Level Orientation
The version of the source container with a beam exit angle of 0° is best suited for
point level detection. Make sure the radiation is directed exactly towards the
detector mounted on the opposite side.
To use larger exit angles (15°, 30°, 45° or 60°), you must certain the beam is
horizontal. To make sure the beam is horizontal, you have to mount the source
holder so that the implied opening of the exit channel is in a horizontal position.
Mount the SHLD 2 as close as possible to the vessel. With large measuring
ranges and small vessel diameters, a gap may exist.
If there are gaps or empty spaces around the installation, provide protective
fences or grids to keep hands away from the dangerous area. You must mark
such areas accordingly.
Figure 4.3
Point Level Orientation
1. Horizontal mounting
2. Vertical mounting
3. Mounting horizontally
,
at right angles to container
Summary of Contents for SHLD 2
Page 1: ...Operating Instructions SHLD 2 Source Holder Document ID 63138 Radiometric...
Page 2: ......
Page 8: ...6 63138 US 200110 NOTES...
Page 12: ...10 63138 US 200110 NOTES...
Page 26: ...24 63138 US 200110 NOTES...
Page 42: ...40 63138 US 200110 NOTES...
Page 56: ...54 63138 US 200110 NOTES...
Page 68: ...66 63138 US 200110 NOTES...
Page 78: ...76 63138 US 200110 9 2 Dimensions Figure 9 7 Source Holder SHLD 2 Standard Version...
Page 79: ...63138 US 200110 77 Figure 9 8 SHLD 2 Beam Angles Figure 9 9 SHLD 2 Height for Source Exchange...
Page 80: ...78 63138 US 200110 Figure 9 10 Source Holder SHLD 2 with Limit Switch...
Page 81: ...63138 US 200110 79 Figure 9 11 Source Holder SHLD 2 with Interlock Safety Switch...
Page 82: ...80 63138 US 200110 Figure 9 12 Source Holder SHLD 2 with Air Actuator...
Page 83: ...63138 US 200110 81 Figure 9 13 Source Holder SHLD 2 with Air Actuator and Limit Switch...
Page 84: ...82 63138 US 200110 Figure 9 14 Source Holder SHLD 2 Explosion Proof with Limit Switch...
Page 86: ...84 63138 US 200110 NOTES...
Page 87: ......