30
63138-US-200110
Density Orientation
Achieve the optimum and most constant conditions for density measurement in
pipes by mounting the measuring equipment on vertical pipelines or conveyors.
Orient the radiation directly towards the detector mounted on the opposite side.
To extend the distance the beam travels through the medium and thus achieve a
better measuring effect, radiate the tube diagonally or use a measuring track.
Find the required mounting accessories in the chapter
"Supplemental-Info"
.
Mount the source holder SHLD 2 as close as possible to the vessel.
However, with large measuring ranges and small vessel diameters, a gap can
often is not avoidable.
If there are gaps or empty spaces around the installation, provide protective
fences or grids to keep hands away from the dangerous area. You must mark
such areas accordingly.
The ideal measurement setup for density measurement is installation on a vertical
pipeline.
Make sure the pipe diameter is at least 1.97 in (50 mm) and the flow direction is
from bottom to top.
Mounting brackets, angled attachments as well as mounting clamps are available
for mounting.
Summary of Contents for SHLD 2
Page 1: ...Operating Instructions SHLD 2 Source Holder Document ID 63138 Radiometric...
Page 2: ......
Page 8: ...6 63138 US 200110 NOTES...
Page 12: ...10 63138 US 200110 NOTES...
Page 26: ...24 63138 US 200110 NOTES...
Page 42: ...40 63138 US 200110 NOTES...
Page 56: ...54 63138 US 200110 NOTES...
Page 68: ...66 63138 US 200110 NOTES...
Page 78: ...76 63138 US 200110 9 2 Dimensions Figure 9 7 Source Holder SHLD 2 Standard Version...
Page 79: ...63138 US 200110 77 Figure 9 8 SHLD 2 Beam Angles Figure 9 9 SHLD 2 Height for Source Exchange...
Page 80: ...78 63138 US 200110 Figure 9 10 Source Holder SHLD 2 with Limit Switch...
Page 81: ...63138 US 200110 79 Figure 9 11 Source Holder SHLD 2 with Interlock Safety Switch...
Page 82: ...80 63138 US 200110 Figure 9 12 Source Holder SHLD 2 with Air Actuator...
Page 83: ...63138 US 200110 81 Figure 9 13 Source Holder SHLD 2 with Air Actuator and Limit Switch...
Page 84: ...82 63138 US 200110 Figure 9 14 Source Holder SHLD 2 Explosion Proof with Limit Switch...
Page 86: ...84 63138 US 200110 NOTES...
Page 87: ......