38
63138-US-200110
Source Interference
When multiple adjacent pipes, vessels, or conveyors have radiometric gauges,
you must consider the orientation of the source beams so each gauge senses
radiation only from its appropriate source. Maximize the distance between the
source holders when mounting them on adjacent vessels. Make sure to maintain
the appropriate source to detector distance as illustrated in
Figure 4.10
.
Figure 4.10
Source/Detector Arrangement for Adjacent Vessels
1. Radiation Beam
2. Detector
3. Source Holders
4.4 Temperature
Your source holder may have components that are sensitive to temperature
extremes. Avoid exposure to temperatures beyond the specified limits. The
maximum ambient temperature is -50 ... +105°C
(-58 ... +221 °F). Consult the
factory if this temperature is exceeded.
Summary of Contents for SHLD 2
Page 1: ...Operating Instructions SHLD 2 Source Holder Document ID 63138 Radiometric...
Page 2: ......
Page 8: ...6 63138 US 200110 NOTES...
Page 12: ...10 63138 US 200110 NOTES...
Page 26: ...24 63138 US 200110 NOTES...
Page 42: ...40 63138 US 200110 NOTES...
Page 56: ...54 63138 US 200110 NOTES...
Page 68: ...66 63138 US 200110 NOTES...
Page 78: ...76 63138 US 200110 9 2 Dimensions Figure 9 7 Source Holder SHLD 2 Standard Version...
Page 79: ...63138 US 200110 77 Figure 9 8 SHLD 2 Beam Angles Figure 9 9 SHLD 2 Height for Source Exchange...
Page 80: ...78 63138 US 200110 Figure 9 10 Source Holder SHLD 2 with Limit Switch...
Page 81: ...63138 US 200110 79 Figure 9 11 Source Holder SHLD 2 with Interlock Safety Switch...
Page 82: ...80 63138 US 200110 Figure 9 12 Source Holder SHLD 2 with Air Actuator...
Page 83: ...63138 US 200110 81 Figure 9 13 Source Holder SHLD 2 with Air Actuator and Limit Switch...
Page 84: ...82 63138 US 200110 Figure 9 14 Source Holder SHLD 2 Explosion Proof with Limit Switch...
Page 86: ...84 63138 US 200110 NOTES...
Page 87: ......