50
63138-US-200110
Use an electrical switching valve, such as the Festo CPE, in the pneumatic line.
With this valve, you can switch off the air supply.
Optionally you can install an additional hand-operated switching valve, such as
the Festo VHEM in the pneumatic line. In emergencies you can then interrupt the
pneumatic air supply on site and switch off the source holder. Mount this hand-
operated switching valve in a safe position outside the radiated area.
The pneumatic actuation can be operated as follows:
1.
Remove the padlock
2.
Pull out the locking bolt.
These steps secure the "OFF" position. The locking bolt is fixed on a steel cable.
3.
Using compressed air, switch the pneumatic actuator from the "OFF"
position to the "ON" position. The pneumatic actuator should move without
interruption to the "ON" position.
Do not grasp any mechanical parts of the pneumatic drive while the
pneumatic actuator is switching over.
Insert the unused locking bolt into its parked position.
Apart from the safety hole, there is another laterally displace hole in
which you can insert the locking bolt for safe retention.
Attach the required padlock to the locking bolt so that the padlock
does not get lost.
!
CAUTION
!
CAUTION
Summary of Contents for SHLD 2
Page 1: ...Operating Instructions SHLD 2 Source Holder Document ID 63138 Radiometric...
Page 2: ......
Page 8: ...6 63138 US 200110 NOTES...
Page 12: ...10 63138 US 200110 NOTES...
Page 26: ...24 63138 US 200110 NOTES...
Page 42: ...40 63138 US 200110 NOTES...
Page 56: ...54 63138 US 200110 NOTES...
Page 68: ...66 63138 US 200110 NOTES...
Page 78: ...76 63138 US 200110 9 2 Dimensions Figure 9 7 Source Holder SHLD 2 Standard Version...
Page 79: ...63138 US 200110 77 Figure 9 8 SHLD 2 Beam Angles Figure 9 9 SHLD 2 Height for Source Exchange...
Page 80: ...78 63138 US 200110 Figure 9 10 Source Holder SHLD 2 with Limit Switch...
Page 81: ...63138 US 200110 79 Figure 9 11 Source Holder SHLD 2 with Interlock Safety Switch...
Page 82: ...80 63138 US 200110 Figure 9 12 Source Holder SHLD 2 with Air Actuator...
Page 83: ...63138 US 200110 81 Figure 9 13 Source Holder SHLD 2 with Air Actuator and Limit Switch...
Page 84: ...82 63138 US 200110 Figure 9 14 Source Holder SHLD 2 Explosion Proof with Limit Switch...
Page 86: ...84 63138 US 200110 NOTES...
Page 87: ......