Corrosion
If there are clear traces of corrosion on the source holder, you must measure the
local dose rate (µSv/h) in the surrounding areas. If the rate is clearly above the
values during normal operation, then you must cordone off the area and inform the
responsible radiation safety officer.
Make certain you exchange the corroded instruments and lock washers as soon
as possible.
Test Shutter Mechanism
Test the function of the shutter mechanism on the source holder at regular
intervals
.
We recommend carrying out this test every six months
.
See the Shutter
(ON/OFF) Mechanism procedure.
Manual Shutter Mechanism
To test the function of the shutter mechanism on a manual shutter, complete the
following steps:
1.
Remove the padlock.
2.
Move the shutter several time from the
"
ON
"
to the
"
OFF
"
position.
If you are not sure of the proper functioning or condition of the
instrument, contact your VEGA representative.
Only the manufacturer. supplier, or specially authorized personnel
may make repairs or maintenance work beyond the scope of
normal inspection
.
Replace source holders that have corroded or stiff locking
devices. Also, replace any source holder that has a shutter
that is not operating correctly.
!
CAUTION
!
CAUTION
!
WARNING
Summary of Contents for SHLD 2
Page 1: ...Operating Instructions SHLD 2 Source Holder Document ID 63138 Radiometric...
Page 2: ......
Page 8: ...6 63138 US 200110 NOTES...
Page 12: ...10 63138 US 200110 NOTES...
Page 26: ...24 63138 US 200110 NOTES...
Page 42: ...40 63138 US 200110 NOTES...
Page 56: ...54 63138 US 200110 NOTES...
Page 68: ...66 63138 US 200110 NOTES...
Page 78: ...76 63138 US 200110 9 2 Dimensions Figure 9 7 Source Holder SHLD 2 Standard Version...
Page 79: ...63138 US 200110 77 Figure 9 8 SHLD 2 Beam Angles Figure 9 9 SHLD 2 Height for Source Exchange...
Page 80: ...78 63138 US 200110 Figure 9 10 Source Holder SHLD 2 with Limit Switch...
Page 81: ...63138 US 200110 79 Figure 9 11 Source Holder SHLD 2 with Interlock Safety Switch...
Page 82: ...80 63138 US 200110 Figure 9 12 Source Holder SHLD 2 with Air Actuator...
Page 83: ...63138 US 200110 81 Figure 9 13 Source Holder SHLD 2 with Air Actuator and Limit Switch...
Page 84: ...82 63138 US 200110 Figure 9 14 Source Holder SHLD 2 Explosion Proof with Limit Switch...
Page 86: ...84 63138 US 200110 NOTES...
Page 87: ......