5 Setup
Electronics module • VEGAFLEX 80 series
43656-EN-170531
After entering the instrument serial number, the order data of the
sensor will be displayed.
3. Below the order data, you can find under "
Sensor data for service
DTM
" the XML file: "DTM Configuration File".
Click with the right mouse key to the file and download this XML
file with "
Save target under
" to your PC.
4. For instruments with SIL qualification you have to download the
PDF document "
SIL electronics exchange check sum certificate
"
also from the homepage.
Click with the right mouse key to the file and download this PDF
file with "
Save target under
" to your PC.
5. Compare the serial number of the XML file with the serial number
of your sensor.
You can find the serial number outside on the type label of the
instrument and inside the housing.
6. Start PACTware and connect to the corresponding sensor.
7. Open the field "
Maintenance
" and select"
Electronics exchange
"
8. After the connection is established, you can choose how the sen-
sor data should be loaded to the sensor.
Choose "
Load sensor data from local file
"
9. Click the button "
Selection
" to load the XML file from the file
system of your computer.
Make sure that you are using the correct file.
10. As soon as the sensor data are ready, you can load the data
directly into the sensor.
This procedure lasts approx. 5 minutes.
11. Then all application-specific settings must be entered again. See
the following chapter "
Setup steps
"
Instruments with SIL qualification
After the transfer of the sensor data, a check sum is displayed. You
have to verify the correct transmission by means of a check sum.
Only then, the instrument will be ready for operation, again and SIL
qualified.
Go in the via our homepage "www.vega.com" to "
Instrument search
(serial number)
".
Enter the serial number of your sensor.
After entering the instrument serial number, the order data of the sen-
sor are displayed.
Download the PDF document "
SIL electronics exchange check sum
certificate
" from the homepage. Compare the check sum with the data
of the PDF document.
If the check sums match, the transmission was successful. Record
the comparison of the check sums.
Summary of Contents for VEGAFLEX FX-E.80H
Page 1: ...Operating Instructions Electronics module VEGAFLEX 80 series Document ID 43656...
Page 17: ...17 Notes Electronics module VEGAFLEX 80 series 43656 EN 170531...
Page 18: ...18 Notes Electronics module VEGAFLEX 80 series 43656 EN 170531...
Page 19: ...19 Notes Electronics module VEGAFLEX 80 series 43656 EN 170531...