17
4 Mounting
VEGAMIP R61 • Receiving unit
35786-EN-171120
1
2 3 4
Fig. 10: VEGAMIP 61, threaded version G1½
1 Polarisation marking
2 Instrument hexagon
3 Counter nut
4 Process seal
1. Insert the supplied process seal (4) before screwing in the instru-
ment
2. To screw VEGAMIP 61 in, use the instrument hexagon (2) below
the housing.
Screw the instrument in and tighten it lightly with a flat spanner
3. Turn the sensor back (max. 180°) to reach the required orientation
of the polarisation marking (1)
4. Hold the instrument in this position and tighten with the counter
nut (3)
Caution:
The housing must not be used to screw the instrument in! Applying
tightening force can damage internal parts of the housing.
Note:
In the case of VEGAMIP 61 with NPT thread, the instrument seals in
the thread itself. Hence, no counter nut is necessary for these ver-
sions.
Avoid long sockets in which the medium can remain and if possible,
mount VEGAMIP 61 front-flush. This applies mainly if buildup and
dust are expected.
Buildup