background image

DL49-MAN    

P. 5

S p e c i f i c a t i o n s

Measurement limits

General specifications

Electrical specifications

Standard & Optional Accessories

Standard

Soft carrying case  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .AC259
Test leads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ATL55

Optional

Plug-in line splitter  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ASL1
Temperature measurement adapter  . . . . . . . . . . . . . . . . . . . . . .TA2K
Hard carrying case (service kits)  . . . . . . . . . . . . . . . . . . . . . . . . .AC506
High voltage probe  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .KHV66
Flame safeguard test adapter  . . . . . . . . . . . . . . . . . . . . . . . . . . .DLMAT

Frequency

AC Amperage
DC Amperage

AC Voltage

DC Voltage

Ohms

Continuity

50 - 400 Hz

400 Amps
400 Amps

600 Volts
600 Volts

40 Megohms

<50 ohms

Size (H x W x L)

Operating temperature

Storage temperature

Relative humidity

(storage and use)

Weight

Calibration frequency

Ul listing

IEC Listing

Battery

Battery life

7.68” x 3.06” x 1.79”

(195 x 77 x 46 mm)

32 to 113˚F (0 to 45˚C)

-4 to 140˚F (-20 to 60˚C)

0% to 80% RH

13 oz. (370g)

Annual

3111-1

1010-1 (CAT III 600 Volts)

9V Alkaline (NEDA 1604

or 6F22 or 006P)

80 hours typical (Alkaline) 

Function

Ohms (

Ω

)

Volts AC

(50-400 Hz)

Volts DC

Amps AC

(50-400 Hz)

Amps DC

Range

400 

Ω

4 K

Ω

40 K

Ω

400 K

Ω

4 MΩ

40 M

Ω

4 V

40 V

400 V
600 V

4V

40 V

400 V
600 V

40 A

400 A

40 A

400 A

Resolution

0.1

Ω

1.0 

Ω

10

Ω

100 

Ω

1 K

Ω

10 K

Ω

1 mV

10 mV

100 mV

1 V

1 mV

10 mV

100 mV

1 V

0.01 A

0.1 A

0.01 A

0.1 A

Accuracy

±(0.9% + 3 digits)

±(1.2% + 5 digits)

±(1.5% + 3 digits)

±(0.9% + 2 digits)

±(2% + 5 digits)

±(3% + 10 digits)

±(2.5% + 10 digits)

±(1.5% + 5 digits)

Reviews: