1 0
::
An inspection log can be found at
point 2.2 in this user manual.
1.8.1. CLEANING
We recommend that the product
and the parts that the user and the
assistant can come into contact
with, are cleaned using a cloth
wringed out with regular cleaning
detergent.
Note:
Strong acids, alkalines and al-
cohol may not be used for cleaning
the VELA Person Lift.
Furthermore, autoclaving may not
be used for cleaning the VELA Per-
son Lift.
The body support
The body support (100% polyester)
must be cleaned using the instruc-
tions found on the body support.
1.8.2. AVOID DEEP DISCHARGE
If the VELA Person Lift is taken out
of operation for a prolonged period,
the emergency stop must be activat-
ed (the red button must be pressed
inwards). This ensures that the battery
is not discharged.
The battery must charge for 24 hours
before the VELA Person Lift is used
again.
1.8.3. PREVENTION OF CORROSION
If the VELA Person Lift gets wet during
use, it should be wiped before stor-
age.
Do not use the body support if it is
damaged.
Check the seams, bands, material,
suspension brackets and the readabili-
ty of the label on the body support.
1.8.4. SERVICE LIFE
The VELA Person Lift has an expected
life service of 10 years, under the pre-
requisite that it is used correctly and
that the annual servicing is done.
The service life depends on how of-
ten the lift is used, and how much the
user weights (max. 125 kg).
The body support is not a part of the
10 year service life.
The body support must be controlled
at least every 6th month by profes-
sionals – see the inspection label on
the body support.
Wash by hand - Do not bleach
Do not iron - Do not dry clean
Do not professional wet clean
Do not tumble dry - Drip line drying