background image

6

Emergency stop

Yes

On/off

Automatically when the VELA Person Lift is used

Protection against overload

Automatic

Protection against deep discharge of battery

Automatic

Power supply

24V DC, 10A

Charger

Input: 100-240V – 50/60Hz,0,5A Output: 28V 0,8A

Battery

24VDC/2.9Ah (corresponding to approximately 50 operation 
cycles depending on conditions)

SWL(safe working load)

125 kg

Duty cycle

Max. 2 min ON/18 min OFF

Charging time

4 hours

Operating temperature

+5°C to +45°C

1.3. SAFETY AND ELECTRONICS

1.3.1. DIMENSIONS

Summary of Contents for Person Lift

Page 1: ...VELA Person Lift VELA Person Lift I N S T R U C T I O N F O R U S E E N Manual no 106953...

Page 2: ...Avoid deep discharge 10 1 8 3 Prevention of corrosion 10 1 8 4 Service life 10 1 9 RECYCLING AND DISPOSAL 11 2 0 INSTALLATION 12 2 0 1 Installation in car 12 2 0 2 Installation on wall 12 2 0 3 Other...

Page 3: ...mportant Please read this user manual carefully and keep it safe for when you need to refer to it This manual is also available on VELA s website www vela eu where you can also find other relevant inf...

Page 4: ...low battery Use Charge 24 h before use Use Charge regularly even when not used to prolong the battery life Use As a lifting device the VELA Person Lift must be subjected to annual inspection and the b...

Page 5: ...ewseconds to reset the lift If it continues to go into overload mode call forservice The VELA Person Lift can only lift with the buttons on the battery and not the hand controller 1 The cable for the...

Page 6: ...attery Automatic Power supply 24V DC 10A Charger Input 100 240V 50 60Hz 0 5A Output 28V 0 8A Battery 24VDC 2 9Ah corresponding to approximately 50 operation cycles depending on conditions SWL safe wor...

Page 7: ...box hand control and actuator is IP54 approved which corresponds to protection against dust and water splashes Labels on the lift should be read at a distance of approx 40 cm under normal daylight co...

Page 8: ...d The assistant must be attentive to the user s well being during lifting Screws that are approved and tested with regards to DS EN 10535 are used Installation and testing of the screws must solely be...

Page 9: ...harger is connected to a regular power grid by the user assistant Charge the battery for 24 hours before first use Button for emergency stop The VELA Person Lift is delivered with the emergency stop a...

Page 10: ...battery is not discharged The battery must charge for 24 hours before the VELA Person Lift is used again 1 8 3 PREVENTION OF CORROSION If the VELA Person Lift gets wet during use it should be wiped b...

Page 11: ...ents and substances can be harmful to human health and to the environment if the waste is not handled correctly Electrical and electronic equipment and batteries are marked with the WEE icon This symb...

Page 12: ...LL The wall bracket must be mounted vertically Fasteners are not included since they depend on the type of wall intended for mounting Attention You may need to obtain permission by the homeowner re ga...

Page 13: ...ody support not being used correctly or due to lack of vigilance from the assistant and or the user 2 0 6 LIFT TO AND FROM A SITTING POSITION In the case of lifting from a manual wheelchair the wheelc...

Page 14: ...fting equipment and the attach ment of body supports should use the VELA Person Lift Plan the relocation and park the car on a flat surface Do not let the user sit in the body support unattended The V...

Page 15: ...1 5...

Page 16: ...s and downwards arrow for moving the lift downwards Operation of the functions require a push force of 4N newton Note Check whether all lifting brackets and straps are placed correctly before the lift...

Page 17: ...um load of 125 kg is exceeded The emergency stop can also be used to minimise standby power when the VELA Person Lift is not used 2 1 4 CHARGING CONNECTION The battery will make a constant beeping sou...

Page 18: ...o the instructions provided by VELA The VELA Person Lift is initially inspected by VELA the week year for the control is stated on the lifts id label Inspection of VELA Person Lift Inspection Notes Da...

Page 19: ...1 9 The body support is inspected before each use and least every 6th month Inspection of body support Inspection Notes Date and signature 01 02 03 04 05 06 07 08 09 10 11 12 13 14 15 16 17 18 19 20...

Page 20: ...701 852 1 Vermund Larsen A S VELA G teborgvej 8 12 9200 Aalborg SV Denmark 45 96 34 76 00 mail vela dk vela eu...

Reviews: