1.3 Symbols 

 

 

Important safety information, refer to the operating manual. 

 

 

Dangerous voltage may be present. 

 

 

Earth ground 

 

 

Double insulation (Protection class II). 

 

1.4 Maintenance 

 

  Before  opening  the  meter,  always  disconnect  test  leads  from  all  sources  of 

electric current. 

  For  continue  protection  against  fire,  replace  fuse  only  with  the  specified 

voltage and current ratings : 

 

     F : F 200mA/250V 

  If  any  faults  or  abnormalities  are  observed,  the  meter  can  not  be  used  any 

more and it has to be checked out. 

  Never use the meter unless the back cover is in place and fastened fully. 

  To  clean  the  meter,  use  a  damp  cloth  and  mild  detergent  only,  do  not  use 

abrasives or solvents on it. 

 

2. Description 

 
This meter is a 4 ½ TRUE RMS professional measuring instrument, capable of 
performing following functions : 
 

- DC and AC voltage measurement 
- DC and AC current measurement 
- Resistance measurement 
- Capacitance measurement 
- Frequency measurement 
- Diode and Transistor test 
- Audible continuity test 

 

 

 

 

Summary of Contents for DVM98

Page 1: ... used with test leads supplied If necessary they must be replaced with the same model or same electric ratings Measuring leads must be in good condition 1 2 During use Never exceed the protection limit values indicated in specifications for each range of measurement When the meter is linked to a measurement circuit do not touch unused terminals When the value scale to be measured is unknown before...

Page 2: ...0V If any faults or abnormalities are observed the meter can not be used any more and it has to be checked out Never use the meter unless the back cover is in place and fastened fully To clean the meter use a damp cloth and mild detergent only do not use abrasives or solvents on it 2 Description This meter is a 4 TRUE RMS professional measuring instrument capable of performing following functions ...

Page 3: ...1 POWER SWITCH 2 DATA HOLD SWITCH 3 CAPACITOR MEASURING SOCKET 4 LCD DISPLAY 5 TRANSISTOR TESTING SOCKET 6 ROTARY SWITCH 7 INPUT JACKS ...

Page 4: ...ce more 2 3 Data Hold Switch When this switch is active the meter will hold the last display reading until pushing the switch to release it 2 4 Input Jacks This meter has four input jacks that are protected against overload to the limits shown During use connect the black test lead to COM jack and connect red test lead depending on the function selected FUNCTION RED LEAD CONNECTION INPUT LIMITS 20...

Page 5: ...00mA current For a maximum of 20A move the red lead to the A jack 2 Set the rotary switch at desired A or A range position and connect test leads in series with the load under measurement The polarity of the red lead connection will be indicated along with the current value when making DC current measurement 3 When only the figure 1 displayed it indicates over range situation and the higher range ...

Page 6: ...mple when measuring 100MΩ resistance the display reading will be 110 00 and the correct measuring result should be 110 00 10 00 100 00MΩ 3 5 Measuring capacitance 1 Set the rotary switch at desired F position 2 Before inserting capacitor under measurement into capacitance testing socket be sure that the capacitor has been discharged fully 3 When measuring capacitors with shorter leads a testing ad...

Page 7: ...nce less than about 50Ω built in buzzer will sound 4 Specifications Accuracy is specified for a period of one year after calibration and at 18 C to 28 C 64 F to 82 F with relative humidity to 80 Accuracy specifications are given as of reading number of least significant digits 4 1 General MAXIMUM VOLTAGE BETWEEN TERMINALS AND EARTH GROUND 1000 V DC or 700 V rms AC sine FUSE PROTECTION mA F 200mA 2...

Page 8: ...s 1 2 of rdg 15 digits Input impedance 2MΩ Frequency Range 40Hz to 1000Hz Response TRUE RMS RMS conversion with Signal Crest Factors Up to 5 4 4 DC Current Range Resolution Accuracy Burden Voltage 2mA 0 1µA 0 5 of rdg 5 digits 110mV mA 20mA 1µA 0 5 of rdg 5 digits 15mV mA 200mA 10µA 0 8 of rdg 5 digits 5 0mV mA 10A 1mA 2 0 of rdg 10 digits 0 03V A 4 5 AC Current Range Resolution Accuracy Burden Vo...

Page 9: ... digits 10 digits Note On 200MΩ range if short input display will read 10MΩ this 10MΩ should be subtracted from measurement results 4 7 Frequency Range Resolution Accuracy 20kHz 1Hz 1 5 of rdg 5 digits Sensitivity 200mV rms and input no more 10V rms 4 8 Capacitance Range Resolution Accuracy 2000pF 0 1pF 20nF 1pF 200nF 10pF 4 0 of rdg 20 digits 2µF 0 1nF 20µF 1nF 5 Accessories Test leads Battery Op...

Page 10: ...hows how to use the holster to 1 Support the meter with a standard angle fig a 2 Support the meter with a small angle using the little stand fig b 3 Hang the meter on the wall using the little stand Take the little stand off from the back side of the large stand and insert it into holes located upper on the holster fig c 4 Hold test leads fig d Fig a Fig b Fig d Fig c ...

Page 11: ...lacement and blow almost always as a result of the operator s error Open the case as mentioned above and then take the PCB out from the front cover Replace the blown fuse with same ratings WARNING Before attempting to open the case be sure that test leads have been disconnected from measurement circuits to avoid electric shock hazard For protection against fire replace fuse only with specified rat...