Display Error Codes
VELOPEX INTRA-X
17
User Guide
Display
Characters
Description
Action
E03
Dryer temperature sensor s/c
Call for Service
E04
Dryer temperature sensor o/c
or connection to Main PCB
Call for Service
E05
Fixer temperature sensor s/c
Call for Service
E06
Fixer temperature sensor o/c
Call for Service
E07
Developer temperature sensor
s/c
Call for Service
E08
Developer temperature sensor
o/c
Call for Service
E10
Fan Sensor Inactive
Call for Service
E11
Drive turns count not zero
Call for Service
E12
Drive motor overloaded
1. Switch off the Machine .
2. Clean the Gears on the
Modules.
3. Switch on the Machine.
4. If still E12 - Call for Service.
E13
Transport motor Sensor
Inactive
or Modules seized
1. Switch off the Machine
2. Clean the gears and belts
on the Modules.
3. Switch on the machine
4. If still E13 - Call for Service
E14
Excessive Fixer / Developer
temperature
Call for Service
E15
Dryer temperature too high
Call for Service
E18
Dryer power duty cycle too low
Call for Service
E33
Watchdog relay test failure
(Low)
Call for Service
E34
Watchdog relay test failure
(High)
Call for Service
tLO
Either Developer or Fixer tank
temperature is low
Wait for the correct
temperature
LId
Lid switch open
Close lid
TR
OUBLE
SHOO
TING