10
Number of High registers
0x00 Low register value
0x00
Number of Low registers
0x03 High register value
0x33
High CRC
0xAD Low register value
0x32
0x00 0x31 0x32 0x33 0x34 0x35
Low CRC
0xFE High register value
0x35
Low register value
0x34
High CRC
0x0A
Low CRC
0x0A
Writing single register: STOP rotation
Request
Field
(Hex)
Description
Address
0x64
Control command
0x06
High start address
0x00
Low start address
0x6A
High register value
0x00
Stop rotation
Low register value
0x01
High CRC
0x61
Low CRC
0xE3
Virtual serial port
Baudrate
9600
Bits
8
Stop Bit
1
Parity
None
Maximum number of
registers for single request
24
6. VELP Ermes Configuration
VELP Ermes is a revolutionary cloud platform that transforms and improves your laboratory experience by creating an ecosystem of
instruments, people and data. The VELP Ermes Cloud Platform is able to reduce distances and accelerate scientific processes in total
safety. In order to access on Ermes, you need to enable your VELP account by selecting "Configure your VELP Ermes account" at
To be able to communicate, the instrument needs to be in the operating range of laboratory Wi-Fi (2.4 GHz) and be configured as follows:
➢
Switch on the CONTROLLER Advance and select AP in the menu Set-up Wi-Fi (see chapter 7.5).
➢
Using the PC/Tablet/Mobile phone, select the RC_SERIAL NUMBER available on the Wi-Fi list, in order to connect directly to the
instrument.
➢
Open a browser on the PC/Tablet/Mobile phone and insert the address 192.168.240.1 to reach the configuration page. Insert
“admin” “admin” when requested as user name and password.
➢
Set the parameters required to connect to your Wi-Fi (network name, password, security, mac address, etc.) according to your
internal procedure and save. If necessary, contact your IT administrator.
➢
Select menu Ermes from the Service menu (see chapter 7.6) and proceed with the product registration from the VELP Ermes Cloud
Platform. For more information see FAQ on VELP website.
NOTE:
To access to VELP Ermes it is necessary to have a VELP account.