35 EN
EMC Information
The appliance is complied with the following emission and immunity standard, and test
level:
Emissions Test Standard
Compliance
RF emissions
CISPR 11
Group 1
RF emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
Comply
Immunity Test standard IEC 60601-1-2 Test Level
Compliance Level
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV
air
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV
air
Electrical fast transient/
burst
IEC 61000-4-4
± 2 kV, for power supply
lines
± 1 kV signal input / output
100 kHz repetition
frequency
± 2 kV, for power supply
lines
± 1 kV signal input / output
100 kHz repetition
frequency
Surge IEC 61000-4-5
± 0.5 kV, ± 1 kV differential
lines
± 0.5 kV, ± 1 kV , ± 2 kV
common mode
± 0.5 kV, ± 1 kV differential
lines
± 0.5 kV, ± 1 kV , ± 2 kV
common mode
Voltage dips, interruptions
and variations on power
supply input lines
IEC 61000-4-11
0 % UT; 0.5 cycle. At 0°,
45°, 90°, 135°, 180°, 225°,
270° and 315°.
0 % UT; 1 cycle and 70 %
UT; 25 / 30 cycles; Single
phase: at 0°.
0 % UT; 250 / 300 cycle
0 % UT; 0.5 cycle. At 0°,
45°, 90°, 135°, 180°, 225°,
270° and 315°.
0 % UT; 1 cycle and 70 %
UT; 25 / 30 cycles; Single
phase: at 0°.
0 % UT; 250 / 300 cycle
Power frequency
(50 / 60 Hz) magnetic field
IEC 61000-4-8
30 A/m
50Hz / 60Hz
30 A/m
50Hz / 60Hz
Conducted RF
IEC 61000-4-6
3 V
0.15 MHz – 80 MHz
6 V in ISM and amateur
radio bands between
0.15 MHz and 80 MHz
80 % AM at 1 kHz
3 V
0.15 MHz – 80 MHz
6 V in ISM and amateur
radio bands between
0.15 MHz and 80 MHz
80 % AM at 1 kHz
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz – 2.7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2.7 GHz
80 % AM at 1 kHz
Note: UT is the AC mains voltage prior to application of the test level.