3.10
RS232 Communication Option
The 301 Controller can be supplied with the capability to communicate with other
devices via an RS232 link. If this option has been ordered, the furnace or oven will be
supplied with a 9 pin ‘D’ socket for connecting to an external device. Plugging this into a
computer will allow the controller to be accessed from that computer. The computer
must have appropriate communication software installed such as Eurotherm's ‘i-Tools’.
RS232 Communication Addressing:
Modbus Address
(Main)
= 2
Modbus Address
(O/Temp)
= 3
Baud Rate
= 9600
Byte Format
= 8
RS232 Communication Cables
Connecting the furnace or oven to a computer is done via a "straight" cable as
follows:
Product end of cable
female 9-pin
RS232 Cable:
product to PC
Computer end of cable
9-pin male
Rx
3
_____________________
3
Tx
Tx
2
_____________________
2
Rx
Com
5
_____________________
5
Com
3.11
Temperature Controller Replacement
Before handling the controller: wear an anti-static wrist strap or
otherwise avoid any possibility of damage to the unit by static electricity.
Refer to the detailed instructions supplied with the replacement
controller.
If the temperature controller is fitted to the back of the control panel it can be
separated from the base by removal of the screws. If the temperature controller is
fitted inside the product base it can be separated from the top by removal of the
screws.
37