12.0 Specifications
12.0
Specifications
Carbolite Gero reserves the right to change the specification without notice.
Model
Max Temp
(°C)
Max Power
(kW)
Chamber Size
(mm)
Approx
Capacity (l)
Net Weight
(kg)
H
W
D
Chamber Furnaces heated by Molybdenum Disilicide elements.
HTF 17/25
1700
12
300
275
300
25
403
12.1
Environment
The models listed in this manual contains electrical parts and should be stored and used
in indoor conditions as follows:
Temperature:
5 °C - 40 °C
Relative
humidity:
Maximum 80 % up to 31 °C decreasing linearly to 50 % at 40
°C
altitude:
not exceeding 2000 m
electrical supply: fluctuation not exceeding 10%
overvoltage:
category II IEC60364-4-443
pollution:
degree 2
62