D.
Chimney Emits Smoke
1.
Some light
smoke is
emitted in the
middle of the
test
No fault
2.
Dense smoke
is emitted
The after-burner
chamber
temperature may
not be set high
enough
Set the secondary controller
temperature to 900 °C
The after-burner
elements or control
may have failed
Investigate or contact Carbolite
Gero
A fuse may have
blown
Check the supply fuses
A Load containing
too much carbon has
been loaded
Reduce the size of the load. set a
much slower heating rate-see
temperature control instructions
49
10.0 Fault Analysis