12.0
Fuses and Power Settings
12.1
Fuses
The following fuses are present:
Supply Fuses: 38 mm x 10 mm type F, 16 A - 6 total
Auxiliary Fuses 20 mm x 5 mm glass type F, 5 A
12.2
Customer Supply Fusing
High break capacity fuses should be used. Avoid fast-blow fuses and magnetic trip
circuit breakers - consult Carbolite Gero if in doubt.
The supply fuse rating should be as follows:
Phases/ Volts
Supply Fuse Rating
3-phase with neutral, 380/ 220 V to 415/ 240 V
16 A per phase
12.3
Power Settings
The furnace control system incorporates electronic power control, including a "power
limit" parameter that is used to reduce the effective voltage to 208 V (or 104 V); the
values of the power limit for different voltages are as follows:
ABF 8/28
Voltage
200 V 208 V
220 V/
380 V
230 V/
400 V
240 V/
415 V
Power (%)
100
92
83
76
69
Please refer to the rating label for product specific information.
51
12.0 Fuses and Power Settings