9.0
Fault Analysis
A.
Furnace Does Not Heat Up
1.
The HEAT
light(s)
are ON.
An ohm meter applied to
the element circuit shows
an open circuit.
A heating element has failed.
2.
The HEAT
light(s)
are OFF.
The controller shows a
very high temperature or
a code such as S.br.
The thermocouple has broken or has
a wiring fault.
The controller shows a
low temperature.
The door switch may be faulty or
need adjustment.
The thyristor fuse may have blown.
The power thyristor could be failing
to switch on due to internal failure,
faulty wiring from the controller, or
fault controller.
There are no lights
glowing on the controller.
Check the supply fuses and any fuses
in the product control compartment.
The controller may be faulty or not
receiving a supply due to a faulty
switch or a wiring fault.