2.6
Gas Connections
The product is fitted with two flow meters, one for Carbon Dioxide (CO
2
) on the left and
the other for Hydrogen (H
2
) on the right. The output from the flow meters is mixed and
taken directly to the work tube through a connection at the front.
There are no valves or non-return devices on this furnace. Regulated
supplies of CO
2
and H
2
must be connected in a safe manner to ensure
that all gas flow is through the work tube. The gas supply pressure
required is 4 psi (0.276 bar) and must not exceed this.
At the back of the furnace is a copper exhaust vent pipe. The exhaust
fumes are dangerous (see "Carbon Monoxide Warning"). The exhaust
must be properly vented into a suitable duct or fume cupboard and must
not be allowed to contaminate the atmosphere of the room in which the
furnace is sited.
11
2.0 Installation