l
Automatic Analysis. Each of the samples is identified by placing a numbered grid
over each individual sample. The operator then has the option to activate the auto-
matic analysis mode. When activated the system will identify the sample profile and
its position within its own grid boundary. The program will then systematically ana-
lyse the deformation of each of the sequential images and identify each of the
deformation points. When the program has finished analysing the images the critical
deformation temperatures are automatically displayed in the results form along
with the individual sample identification. Graphical data of the melting profiles are
also recorded. The results table can then be saved and printed out in report form.
Note
: Only 8 samples can be automatically analysed at any one time.
l
Manual Analysis. To aid analysis each of the samples is identified by placing a
numbered grid over it. Analyses is aided by the use of a results form, which enables
the critical deformation temperatures to be entered manually along with individual
sample identification. The results form can then be saved and printed out in report
form.
In both cases the software automatically starts and stops the furnace controller, the
start temperature and interval between images is adjustable to suit your requirements.
See section 4.7.2.2.
The software is designed primarily to be used with a mouse, however most operations
are selectable from the keyboard if necessary and are as follows:
l
Commands on the text menu bar at the top left of the screen are activated using the
'Alt' key together with the first letter of the respective command. For drop down
menus select the first letter of the respective command without using the 'Alt' key.
l
When using the 'play back' facility the images can be scrolled back and forth by drag-
ging the scroll bar, or using the left and right 'arrow' keys. When the results window
is active the < and > keys will scroll the images.
Language Options
Initial language settings can be set by choosing from a list of 11 languages, by
selecting 'Tools', 'Languages'.
Chinese
English
Japanese
Czech
French
Russian
Danish
German
Spanish
Dutch
Hungarian
27
Summary of Contents for CARBOLITE GERO CAF G5
Page 63: ...Fig 3 Front Tube Seal Assembly Fig 4 Door Arm Assembly 63 ...
Page 64: ...Fig 5a Camera Mounting Bracket Fig 5b Lens and Camera Assembly 64 ...
Page 65: ...Fig 5c Sliding the Camera Mounting Bracket Assembly onto the Door Arm 65 ...
Page 66: ...Fig 5d Securing the Camera Mounting Bracket Assembly to the Door Arm 66 ...
Page 67: ...Fig 5e Mounting the Lens and Camera Assembly 67 ...
Page 68: ...Fig 5f Lens and Camera Assembly in Position 68 ...
Page 69: ...Fig 6a Rear View of the Standard CAF G5 Furnace Showing Brick Box Assembly 69 ...
Page 71: ...Fig 7 Front Tube Seal Position 71 ...
Page 72: ...Fig 8 Tube End Seal Assembly Tightening Sequence Fig 9 Work Tube Front Support 72 ...
Page 73: ...Fig 10 Fitting the Door Arm Assembly Fig 10a Adjusting the Door Arm Assembly 73 ...
Page 74: ...Fig 11a Furnace Case and Controls 74 ...
Page 75: ...Fig 11b Gas Inlet Pipe Fig 12 Positioning the Furnace 75 ...
Page 77: ...Fig 13c Loading Samples into the Work Tube 77 ...
Page 78: ...Fig 16 File Folder 78 ...
Page 79: ...Fig 17 Door Arm Assembly Exploded View 79 ...
Page 82: ...SST DT HT FT Fig 18 Report Sheet Page 2 Side View Plan View Fig 19 Formed Wire Sample 82 ...
Page 83: ...Fig 20 Sample Carrier Sample Tiles and Sample Positions 83 ...
Page 84: ...Fig 21a Coal and Coke Test Piece Mould Fig 21b Biomass Test Piece Mould and Hand Press 84 ...
Page 85: ...Fig 22 Sample Loading Tool Fig 23 Camera Ethernet Connection 85 ...
Page 86: ...Fig 24 LED Driver Connection 86 ...
Page 87: ...Notes Service Record Engineer Name Date Record of Work ...