Sample Reference Information
- The upper part of this area shows the test details
such as 'Test Name', 'Test Reference', 'Acquisition date/time' and 'Analyse Date/ Time'.
The lower section controls the test piece identity number and the image number and/ or
temperature of the 'Current Selected Temperature Point' (this is shown by a green line
headed 'Current').
To show the parameters of a test piece enter the number of the test piece into the 'Test
Piece' box. This will highlight the corresponding sample grid in the sample window and
show the parameter data and the graphical data for the selected sample.
The current selected temperature point can be advanced by entering the image number
into the 'Image' box or by selecting a defined temperature point. This will also advance
the sample image in the Sample Window.
Parameter Selection
- The parameters for each individual test piece can be shown on
the graph by selecting the tick box in the graph column. Individual or all parameters can
be selected at any time. The graph key information will be shown at the right hand side
of the graph when a parameter is selected.
Deformation Selection
- The deformation temperature points can be shown on the
graph by selecting the tick box in the 'Marker ON' row under each of the deformation
point columns. Each deformation point when selected is highlighted with a magenta
coloured line which is headed with its deformation identification i.e. SST, DT, HT, FT.
Analysis Graph
- The graph shows any parameters or deformation points that have
been selected from the Parameter selection or the Deformation selections.
Note
: No graph will be shown if none of the parameters are selected.
4.7.4.9
Manual Entry of Analysis Results
To enter results manually, the manual analysis tab should be first selected.
Note
: Automatic and manual results are saved independently of each other in two
separate folders, see section 4.7.4.10.
The results table should be used in conjunction with the sample grids therefore the grids
should be activated prior to opening the Results Form, see section 4.7.2.4.
To enter the selected deformation temperature for a test piece into the form, the
sample number must first be selected. The active test piece is then identified by a
highlighted blue row on the form.
The test piece number row can be activated in one of two ways:
l
Selecting a corresponding grid in the test window, i.e. if grid 2 is selected (this will be
highlighted red), row 2 on the form will be highlighted in blue.
l
Selecting the 'Active Grid' number by scrolling up and down using the up/ down
arrows or by entering a number directly into the 'Active Grid' box.
Once the test piece row has been selected the identified deformation temperatures can
be entered into the relevant columns, by simply selecting the cell which the temperature
is to be entered.
To find the relevant deformation image, the test images can be scrolled through in a
number of ways:
42
Summary of Contents for CARBOLITE GERO CAF G5
Page 63: ...Fig 3 Front Tube Seal Assembly Fig 4 Door Arm Assembly 63 ...
Page 64: ...Fig 5a Camera Mounting Bracket Fig 5b Lens and Camera Assembly 64 ...
Page 65: ...Fig 5c Sliding the Camera Mounting Bracket Assembly onto the Door Arm 65 ...
Page 66: ...Fig 5d Securing the Camera Mounting Bracket Assembly to the Door Arm 66 ...
Page 67: ...Fig 5e Mounting the Lens and Camera Assembly 67 ...
Page 68: ...Fig 5f Lens and Camera Assembly in Position 68 ...
Page 69: ...Fig 6a Rear View of the Standard CAF G5 Furnace Showing Brick Box Assembly 69 ...
Page 71: ...Fig 7 Front Tube Seal Position 71 ...
Page 72: ...Fig 8 Tube End Seal Assembly Tightening Sequence Fig 9 Work Tube Front Support 72 ...
Page 73: ...Fig 10 Fitting the Door Arm Assembly Fig 10a Adjusting the Door Arm Assembly 73 ...
Page 74: ...Fig 11a Furnace Case and Controls 74 ...
Page 75: ...Fig 11b Gas Inlet Pipe Fig 12 Positioning the Furnace 75 ...
Page 77: ...Fig 13c Loading Samples into the Work Tube 77 ...
Page 78: ...Fig 16 File Folder 78 ...
Page 79: ...Fig 17 Door Arm Assembly Exploded View 79 ...
Page 82: ...SST DT HT FT Fig 18 Report Sheet Page 2 Side View Plan View Fig 19 Formed Wire Sample 82 ...
Page 83: ...Fig 20 Sample Carrier Sample Tiles and Sample Positions 83 ...
Page 84: ...Fig 21a Coal and Coke Test Piece Mould Fig 21b Biomass Test Piece Mould and Hand Press 84 ...
Page 85: ...Fig 22 Sample Loading Tool Fig 23 Camera Ethernet Connection 85 ...
Page 86: ...Fig 24 LED Driver Connection 86 ...
Page 87: ...Notes Service Record Engineer Name Date Record of Work ...