5.0 Operation
maximum, or heat using a slowly controlled ramp rate. For more information refer to
the controller instructions.
When heating materials that produce smoke or fumes, the chimney must be correctly
fitted and unobstructed. If not, soot will accumulate in the chamber and could possibly
cause an electrical breakdown of the heating element. If the furnace is used to heat
materials that emit smoke or fumes, regularly heat it up to maximum temperature for
one hour with the chamber empty to burn away the soot.
5.3
Atmospheres
When an optional gas inlet is fitted, there is a label near the inlet saying "INERT GAS
ONLY". In practice, inert or oxidising gases may be used, but not combustible or toxic
gases.
The chamber is not gas tight, the gas usage may be high and the chamber is always
likely to contain some air. Residual oxygen of approximately 1% to 2% is to be
expected.
5.4
Operator Safety
This product has mineral insulated heating elements and does not require a door safety
switch.
Avoid burns. Carbolite Gero can supply tongs, face masks and heat resistant gloves.
Before you remove a hot object from the product, make sure you have a safe place to
put it down.