7.7
Power Adjustment (Thyristor)
The current-limiting thyristor stacks which control power to the elements are fitted with
an adjustable resistor which is factory set to limit the maximum current supplied. In the
event of a change of supply voltage, or the fitting of a new thyristor, further adjustment
may be required.
The maximum element currents for this model are listed in section 11.0. Please contact
Carbolite Gero for further information.
7.8
Low Voltage Compensation
If the supply voltage proves to be routinely below the nominal figure for which the
furnace has been set up, there is a "+2%" position on the transformer primary side
which may be used instead of the standard position, see section 8.10.
51
7.0 Maintenance
Summary of Contents for CARBOLITE GERO HTF 18/8
Page 7: ...Caution Double Pole Neutral Fusing 7 1 0 Symbols and Warnings ...
Page 37: ...Power Limit View 37 4 0 3508 Dual Loop Cascade ...
Page 59: ...59 8 0 Repairs and Replacements ...
Page 64: ...10 0 Wiring Diagrams GR Y Green Yellow 64 ...
Page 68: ......
Page 69: ...Notes Service Record Engineer Name Date Record of Work ...