8.0 Repairs and Replacements
In the case of the 380-415 V version,
there are three tappings marked 380,
400 and 415.
The thermal trip connection forms part of the circuit to the contactor coil.
The L and N connection may alternatively be L1 and L2 for a live-to-live supply. The L
side should be connected to the correct tapping to match the electrical supply voltage.
The link wire should be in position as shown unless the electrical supply voltage is always
low, in which case there is the option of moving one end to the +2% terminal.
Important - Changing the transformer primary tapping (including the +2%) requires
adjustment of the thyristor current - see section 8.4.
The terminal between the link connection is not used.
8.11
Fuse Replacement
Fuses are accessed by removal of the appropriate panel, as described in the 'Panel
Removal' section. Depending on the model, supply fuses and control circuit fuses may
be mounted in their own holders, or may be on a circuit board that contains an EMC
filter. The fuses are marked with their ratings.
Take care not to disconnect the wires leading from the EMC filter without first recording
their positions: they must be reconnected to the correct terminals.
Note:
The main fuses for PF 30 models are located in an external fuse holder
positioned next to the IEC socket for the power supply cable.
58
Summary of Contents for CARBOLITE GERO HTF 18/8
Page 7: ...Caution Double Pole Neutral Fusing 7 1 0 Symbols and Warnings ...
Page 37: ...Power Limit View 37 4 0 3508 Dual Loop Cascade ...
Page 59: ...59 8 0 Repairs and Replacements ...
Page 64: ...10 0 Wiring Diagrams GR Y Green Yellow 64 ...
Page 68: ......
Page 69: ...Notes Service Record Engineer Name Date Record of Work ...