6.0 Repairs and Replacements
Remove the thermocouple(s) by withdrawing them from the sheaths in the elements.
Remove the plates through which the element tails are located. Remove the keep plates
from each side of the insulation assembly.
Lift out the element to be replaced; save any insulation sleeves for possible reuse.
Bend or cut the new element tails as necessary and fit any insulation sleeves; feed the
tails through and fit the element into place.
Refit the keep plates on each side. Refit the tail termination plates, ensuring that the
element tails do not touch any metal parts. Refit the thermocouple(s).
Connect all the wiring according to the plan previously made and complete the
reassembly of the product.
Check that the product is controlling properly to rule out the possibility that the element
failed because of a fault in the control system.
6.7
Fuse Replacement
Fuses are marked on the wiring diagram with type codes, e.g. F1, F2. For more
information on fuses refer to section 9.0.
Depending on model and voltage, the different fuse types may or may not be fitted.
If any fuse has failed, it is advisable for an electrician to check the internal circuits.
Replace any failed fuses with the correct type. For safety reasons do not fit larger
capacity fuses without first consulting Carbolite Gero.
The fuses are located at the cable entry point. Remove the back panel or control box
back panel to gain access to the fuses.
24